{"id":"ff5057e9-c668-4fb7-9229-52a4bd005448","slug":"spinal-cord-stimulator-failure-migration-of-a-thoracic-epidural-paddle-to-the-cervical-spine","title":"Spinal cord stimulator failure: Migration of a thoracic epidural paddle to the cervical spine","authors":["Sarah S. Travers","Thorkild V. Norregaard"],"abstract":"Background: Spinal cord stimulators successfully treat a number of pain syndromes but carry a risk of hardware complications. Here, we present a case of cranial migration of a thoracic epidural paddle to the cervical spine. Case Description: A 53-year-old male underwent uncomplicated spinal cord stimulator placement at the T10– T11 with initially favorable results. However, postoperatively, he complained of paresthesias in his arms. An X-ray demonstrated cranial migration of the thoracic epidural paddle to the cervical spine. The stimulator/new paddle was placed again at the T10–T11 level, but the leads were now secured to the caudal lamina utilizing a cranial plating system. The patient subsequently did well without further sequelae. Conclusions: A thoracic epidural paddle (T10–T11) migrated postoperatively into the cervical spine. It was subsequently removed and replaced into the thoracic region, but the leads were now secured in place with a novel caudal lamina/cranial plating system.","thumbnailUrl":"https://sni-digital-videos.s3.amazonaws.com/articles/ff5057e9-c668-4fb7-9229-52a4bd005448/featured/hero-1781563881656.png","publishDate":"2019-06-25T00:00:00.000Z","doi":"10.25259/SNI-302-2019","categories":["Unique Case Observations","Case Report"],"fullTextUrl":"https://surgicalneurologyint.com/api/articles/ff5057e9-c668-4fb7-9229-52a4bd005448/pdf"}