Tools

Shinya Kobayashi, Tatsuya Ishikawa, Tatsushi Mutoh, Kentaro Hikichi, Akifumi Suzuki
  1. Department of Surgical Neurology, Research Institute for Brain and Blood Vessels-AKITA, Akita, Japan

Correspondence Address:
Shinya Kobayashi
Department of Surgical Neurology, Research Institute for Brain and Blood Vessels-AKITA, Akita, Japan

DOI:10.4103/2152-7806.102330

Copyright: © 2012 Kobayashi S. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

How to cite this article: Kobayashi S, Ishikawa T, Mutoh T, Hikichi K, Suzuki A. A novel technique for ventriculoperitoneal shunting by flat panel detector CT-guided real-time fluoroscopy. Surg Neurol Int 13-Oct-2012;3:119

How to cite this URL: Kobayashi S, Ishikawa T, Mutoh T, Hikichi K, Suzuki A. A novel technique for ventriculoperitoneal shunting by flat panel detector CT-guided real-time fluoroscopy. Surg Neurol Int 13-Oct-2012;3:119. Available from: http://sni.wpengine.com/surgicalint_articles/a-novel-technique-for-ventriculoperitoneal-shunting-by-flat-panel-detector-ct-guided-real-time-fluoroscopy/

Date of Submission
14-Jul-2012

Date of Acceptance
27-Aug-2012

Date of Web Publication
13-Oct-2012

Abstract

Background:Surgical placement of a ventriculoperitoneal shunt (VPS) is the main strategy to manage hydrocephalus. However, the failure rate associated with placement of ventricular catheters remains high.

Methods:A hybrid operating room, equipped with a flat-panel detector digital subtraction angiography system containing C-arm cone-beam computed tomography (CB-CT) imaging, has recently been developed and utilized to assist neurosurgical procedures. We have developed a novel technique using intraoperative fluoroscopy and a C-arm CB-CT system to facilitate accurate placement of a VPS.

Results:Using this novel technique, 39 consecutive ventricular catheters were placed accurately, and no ventricular catheter failures were experienced during the follow-up period. Only two patients experienced obstruction of the VPS, both of which occurred in the extracranial portion of the shunt system.

Conclusion:Surgical placement of a VPS assisted by flat panel detector CT-guided real-time fluoroscopy enabled accurate placement of ventricular catheters and was associated with a decreased need for shunt revision.

Keywords: Cone-beam computed tomography, hydrocephalus, shunt failure, ventriculoperitoneal shunt

INTRODUCTION

Surgical placement of a ventriculoperitoneal shunt (VPS) is the main strategy to manage hydrocephalus. However, the failure rate associated with placement of ventricular catheters remains high. Ventricular catheter placement traditionally involves the identification of external landmarks to determine the catheter entry point and a blind pass via typical trajectories, with the hope that the placement of the catheter is adequate. However, the failure rate of ventricular catheter systems remains as high as 30–40% in the first year,[ 3 7 ] leading to a high incidence of shunt removal or revision.[ 9 ] Those failures are most commonly related to the proximal occlusion of the shunt catheters.[ 6 13 ]

The advantage of the posterior approach for VPS placement is the ease of tunneling via a straight path down to the abdomen.[ 3 10 ] In contrast, the frontal approach requires an additional curved path with a separate incision. However, optimal ventricular cannulation from the posterior perspective is more challenging when compared with anterior placement,[ 11 ] and the choroid plexus often obstructs the holes located at the tip of the ventricular catheter, especially when the catheter is inserted through the posterior approach. Thus, to prevent potential obstruction by the choroid plexus, the tip of the catheter should not be in proximity to the choroid plexus. In recent years, the use of endoscopic[ 8 14 ] and navigation[ 1 5 ] assistance has increased the ability to more precisely place shunts, although each of these techniques has its own advantages and disadvantages.

The use of a hybrid operating room (OR) equipped with radiological examination modalities can facilitate neurosurgical procedures. For example, such systems have enabled intraoperative evaluation with two-/three-dimensional (2D/3D) angiography, fluoroscopic imaging, and soft-tissue cross-sectional imaging,[ 12 ] thereby improving the efficacy and safety of surgical procedures.

Our institution has utilized C-arm-based real-time fluoroscopy and cone-beam computed tomography (CB-CT) in an effort to facilitate a safer and more accurate ventricular catheter placement. This study describes our 3-year experience using this new technique for VPS surgery.

MATERIALS AND METHODS

Patient population

Thirty-nine patients (11 male and 28 female; age range, 44–80 years; mean age, 65.3 years) who underwent VPS surgery with the new technique between June 2008 and May 2011 were enrolled. VPS surgery was performed for 36 patients with normal pressure hydrocephalus after subarachnoid hemorrhage, for one patient with idiopathic normal pressure hydrocephalus, and for two patients with obstructive hydrocephalus (in one patient after surgery for angioblastoma and another patient after intracerebral hemorrhage; Table 1 ).


Table 1

Clinical summary of this novel ventriculoperioneal shunting

 

As a comparison, we analyzed 37 patients (13 male and 24 female; age range, 21–83 years; mean age, 64.4 years) who underwent VPS surgery via conventional methods using the standard external landmarks[ 2 4 ] between January 2002 and May 2008 [ Table 2 ].


Table 2

Clinical summary of conventional ventriculoperioneal shunting

 

The surgical outcomes and the rate of complications were compared between the two patient groups.

Novel surgical technique

Our hybrid OR has a newly designed, multipurpose radiolucent surgical table with a special radiolucent head clamp system and a digital subtraction angiography (DSA) system with a biplane C-arm. For image viewing, there are seven flat-display monitors equipped for biplane angiographic imaging and 3D imaging. During surgery, the monitor can be easily repositioned to assist visualization. In addition to the apparatus for conventional 2D and 3D DSA, a newly developed C-arm CB-CT imaging system (Dyna CT, Siemens, Germany) has been installed. This advanced device provides bone and soft-tissue images. The high-resolution 3D image data set was reconstructed using the OR 3D Workstation.[ 12 ] For 3D image acquisition, the anteroposterior C-arm moved continuously over 220° in 20 seconds.

Surgery was performed with patients under general anesthesia. Patients were placed in this supine position with his/her head rotated to the contralateral side, and the neck was extended with a roll placed under the shoulder. A carbon head clamp adapter and a carbon Mayfield head clamp were used. To reduce artifacts made when C-arm CB-CT imaging was performed, the position of the head pins and the spring of the Mayfield head clamp were kept as far as possible from the expected planes for ventricular catheter placement.

First, copper markers were placed on the glabella and on the conventional entry point of the posterior approach (6 cm above and 3 cm lateral to the inion) as external landmarks, and C-arm CB-CT imaging was performed to provide reconstructed axial, sagittal, and coronal images. The actual point for burr hole, the direction for catheter insertion, and the depth of the catheter was determined based on these images [ Figure 1a ]. External copper markers placed on the glabella and the conventional entry point were detected with CB-CT to allow correction of the true external landmark relative to the external copper maker based on visual inspection [ Figure 1b ].


Figure 1

(a) Planning CB-CT produces three-dimensional surgical planes (axial, coronal, and sagittal) and a three-dimensionally reconstructed image, which helps determining the site of the burr hole and the direction and depth for catheter insertion. (b) External copper markers placed on the glabella and the conventional entry point can be detected with planning CB-CT to allow correction of the true external landmark relative to the external copper maker based on visual inspection

 

Next, the surgical field was disinfected and draped, and a burr hole was made in the appropriate position. Under fluoroscopic imaging, the ventricle was punctured along the planned trajectory until the tip of the ventricular catheter reached the target position in the anterior horn. Visual corrections regarding the direction of the puncture and catheter insertion were made using a virtual plane constructed by the points of puncture and target as well as the plane parallel to the flat-panel detector [ Figure 2 ]. To prevent obstruction by the choroid plexus, the tip of the catheter was placed 2 cm deep from the foramen of Monro, as the ventricular catheter employed in this procedure had side holes within 1.8 cm from the tip.


Figure 2

(a) Virtually constructed planes (a and b) cover the points of puncture and the expected final position of the catheter tip. Plane (a) is consistent with the fluoroscopic image. Plane (b) is vertical to plane. (a) The intersection of the planes indicates the appropriate direction for catheter insertion (arrow). (b) Fluoroscopic image during catheter insertion. Accurate direction and depth for catheter insertion (yellow line) is overlaid on the fluoroscopic image. (c) Intraoperative image. The operator punctures the ventricle, referring to the fluoroscopic guidance

 

Finally, C-arm CB-CT imaging was performed to confirm that the ventricular catheter was situated in an appropriate position [ Figure 3 ]. If necessary, fine adjustment was made under a fluoroscope. Then, all shunt tubes were connected together, and the peritoneal catheter was placed in the abdominal cavity; its positioning was also confirmed using fluoroscopy.


Figure 3

Reconstructed CB-CT images after the catheter insertion. These images allowing confirmation of ventricular catheter positioning. The catheter is inserted along the planned trajectory (red, blue, and green lines), and the tip reaches the target point

 

RESULTS

Among the 37 patients undergoing the conventional VPS technique, 11 (29.7%) ventricular catheters were placed inappropriately; 9 were placed in the contralateral anterior horn, 1 was placed in the ipsilateral inferior horn, and 1 was placed in the prepontine cistern. In the follow-up period described (mean 28 months, median 12 months, range 1–81months), four (10.9%) revision surgeries have been necessary; two ventricular catheters required revision because of misplacement or proximal obstruction, one peritoneal catheter was revised because of distal obstruction; and one entire shunt system was removed because of infection [ Table 2 ].

Using the new technique, all 39 consecutive ventricular catheters were placed accurately, with the tip of each catheter placed at the target point in the ipsilateral anterior horn. We have not experienced any ventricular catheter failures during the follow-up period (mean 14 months, median 12 months, range 1–48 months). Two (5.1%) patients developed shunt obstruction in the extracranial portion of the shunt system; one required revision of the peritoneal catheter, and the other required reconnection of the shunt valve to the peritoneal catheter [ Table 1 ].

Representative case

A 57-year-old woman developed normal pressure hydrocephalus after subarachnoid hemorrhage. She underwent VPS surgery, in which the ventricular catheter was inserted through the right occipital region using the technique described. The tip of the ventricular catheter was placed at the exact target in the anterior horn, and the VPS system has worked well [ Figure 4 ].


Figure 4

(a) CT scan of a 57-year-old woman who developed normal pressure hydrocephalus after subarachnoid hemorrhage is shown. The ventricles are enlarged and are associated with periventricular lucency. (b) CT after surgery. The tip of the ventricular catheter was placed at the exact target in the anterior horn, and ventricle size returned to normal

 

DISCUSSION

The complication rate of VPS surgery ranges from 30% to 40%,[ 3 7 ] with shunt infection and malfunction of the ventricular catheter among the most frequent types of major complications.[ 6 13 ] Blind insertion of ventricular catheter using an external landmark as a reference is associated with suboptimal positioning and an increased risk of complications. Indeed, as many as 29.7% of our patients undergoing conventional VPS placement experienced inappropriate positioning of the ventricular catheter, and two of these patients required repeat surgery for shunt revision.

Endoscopic- and navigation-assisted shunt techniques have been used to facilitate more accurate shunt placement. Villavicencio et al.[ 14 ] compared the revision rates of VPS systems in endoscopically and nonendoscopically placed ventricular shunt catheters and concluded that proximal failure was less likely with optimally placed catheters. In contrast, the endoscopic shunt insertion trial reported that the shunt failure rate were not significantly different when comparing the endoscopic and control groups at 1 year after the surgery.[ 8 ] In the latter study, placement of the tip of the ventricular catheter away from the choroid plexus did not differ significantly between groups. Moreover, even in the endoscopic shunt insertion technique, the position of the burr hole and the catheter trajectory were based on anatomical landmarks. As a result, inappropriate trajectory of catheter may have affected the final ventricular catheter position.

In terms of navigation-assisted techniques, Azeem et al.[ 1 ] investigated 34 ventricular catheters that were placed using the Medtronic electromagnetic frameless neuronavigation system and suggested that this method was helpful for accurate insertion of ventricular catheters. Another multicenter prospective cohort study that compared navigated shunt placement with standard blind shunt placement concluded that electromagnetic-navigated shunt placement reduced shunt misplacement, resulting in a significant decrease in the early shunt revision rate.[ 5 ] However, this technology cannot generate real-time information to detect any collapse of the ventricles with the loss of cerebrospinal fluid. Moreover, this tool cannot show the position of the tip of the ventricular catheter intraoperatively.

The present study demonstrated that accurate catheter placement was associated with a decrease in the incidence of proximal shunt failure, although this study had a relatively short follow-up period. This study also demonstrated that VPS placement under flat panel detector CT-guided real-time fluoroscopy enabled accurate catheter placement. In addition, this technique makes it possible to reveal and arrange the position of the ventricular catheter intraoperatively. The patient receives some amount of irradiation through CB-CT, but the amount of radiation associated with CB-CT for VPS surgery is similar to that associated with a single CT scan, which is within the safe range for adult patients. Artifacts from pins and coils in the Mayfield head clamp can compromise the quality of the CB-CT image; thus, modifications of the materials may be necessary to improve the diagnostic value of this modality.

CONCLUSIONS

The present study demonstrated the utility of flat panel detector CT-guided real-time fluoroscopy for accurate placement of ventricular catheters during VPS surgery. Accurate shunt placement with this novel technique may help reduce the early shunt revision rate.

ACKNOWLEDGEMENTS

We thank Kaoru Sato and Keita Narita for their invaluable support in the acquisition of radiological and intraoperative data.

References

1. Azeem SS, Origitano TC. Ventricular catheter placement with a frameless neuronavigational system: A 1-year experience. Neurosurgery. 2007. 60: S243-7

2. Black PM, Youmans JR.editors. Hydrocephalus in adults. Neurological Surgery. Philadelphia: WB Saunders; 1996. p. 927-44

3. Drake JM, Kestle JR, Milner R, Cinalli G, Boop F, Piatt J. Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus. Neurosurgery. 1998. 43: 294-303

4. Greenberg MS.editors. CSF diversionary procedures. Handbook of Neurosurgery. New York: Thieme; 2010. p. 207-14

5. Hayhurst C, Beems T, Jenkinson MD, Byrne P, Clark S, Kandasamy J. Effect of electromagnetic-navigated shunt placement on failure rates: A prospective multicenter study. J Neurosurg. 2010. 113: 1273-8

6. Kang JK, Lee IW. Long-term follow-up of shunting therapy. Childs Nerv Syst. 1999. 15: 711-7

7. Kestle J, Drake J, Milner R, Sainte-Rose C, Cinalli G, Boop F. Long-term follow-up data from the Shunt Design Trial. Pediatr Neurosurg. 2000. 33: 230-6

8. Kestle JR, Drake JM, Cochrane DD, Milner R, Walker ML, Abbott R. Lack of benefit of endoscopic ventriculoperitoneal shunt insertion: A multicenter randomized trial. J Neurosurg. 2003. 98: 284-90

9. Korinek AM, Fulla-Oller L, Boch AL, Golmard JL, Hadiji B, Puybasset L. Morbidity of ventricular cerebrospinal fluid shunt surgery in adults: An 8-year study. Neurosurgery. 2011. 68: 985-94

10. Li V, Dias MS. The results of a practice survey on the management of patients with shunted hydrocephalus. Pediatr Neurosurg. 1999. 30: 288-95

11. Lind CR, Tsai AM, Law AJ, Lau H, Muthiah K. Ventricular catheter trajectories from traditional shunt approaches: A morphometric study in adults with hydrocephalus. J Neurosurg. 2008. 108: 930-3

12. Murayama Y, Saguchi T, Ishibashi T, Ebara M, Takao H, Irie K. Endovascular operating suite: Future directions for treating neurovascular disease. J Neurosurg. 2006. 104: 925-30

13. Sainte-Rose C, Piatt JH, Renier D, Pierre-Kahn A, Hirsch JF, Hoffman HJ. Mechanical complications in shunts. Pediatr Neurosurg. 1991. 17: 2-9

14. Villavicencio AT, Leveque JC, McGirt MJ, Hopkins JS, Fuchs HE, George TM. Comparison of revision rates following endoscopically versus nonendoscopically placed ventricular shunt catheters. Surg Neurol. 2003. 59: 375-9

Leave a Reply

Your email address will not be published. Required fields are marked *