- Division of Pediatric Neurosurgery, CHOC Children’s Neuroscience Institute, California, United States
- Department of Research, CHOC Children’s Research Institute, California, United States
- Division of Pediatric Neurosurgery, CHOC Children’s Hospital, Orange, California, United States
Correspondence Address:
Leandro Castaneyra-Ruiz, Department of Research, CHOC Children’s Research Institute, Orange, California, United States.
DOI:10.25259/SNI_970_2024
Copyright: © 2025 Surgical Neurology International This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-Share Alike 4.0 License, which allows others to remix, transform, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.How to cite this article: Bianca Romero1, Gio Jison2, Scott Self3, Seunghyun Lee2, Sora Sato2, Celine Thao-Quyen Tran2, Leandro Castaneyra-Ruiz2, Michael Muhonen3. Absence of immunoreaction and cellular adhesion in a polyvinylpyrrolidone-coated ventricular catheter with choroid plexus obstruction: A case report. 28-Feb-2025;16:65
How to cite this URL: Bianca Romero1, Gio Jison2, Scott Self3, Seunghyun Lee2, Sora Sato2, Celine Thao-Quyen Tran2, Leandro Castaneyra-Ruiz2, Michael Muhonen3. Absence of immunoreaction and cellular adhesion in a polyvinylpyrrolidone-coated ventricular catheter with choroid plexus obstruction: A case report. 28-Feb-2025;16:65. Available from: https://surgicalneurologyint.com/?post_type=surgicalint_articles&p=13421
Abstract
BackgroundWhile a variety of modalities are available for the treatment of hydrocephalus, ventriculoperitoneal shunting (VPS) remains the most utilized treatment. Although efficacious, VPS is susceptible to malfunction, with catheter obstruction as the primary cause of failure in pediatric patients. Prior studies have speculated that implanted catheters trigger an immune response from the central nervous system, resulting in cellular reactivity and subsequent obstruction of the device. These cells are derived from the choroid plexus (ChP), which plays an active role in immunological surveillance. Its cellular components contain some of the putative cells that contribute to ventricular catheter occlusion.
Case DescriptionThe case illustrated herein is a patient with a functionally obstructed polyvinylpyrrolidone (PVP)-coated catheter, with ChP occluding the catheter fenestrations. While silicone catheter obstruction typically presents with fibrosis and microglial reaction, the illustrated case demonstrates the absence of an immunological response. PVP-coated catheters appear to deter cellular attachment which may dampen the immune response to the catheter in the brain. However, the case discussed postulates that ChP can still obstruct PVP-coated catheters through growth and expansion into the catheter holes and lumen, even without an immune response.
ConclusionThis case report highlights the complexity of novel catheter designs constructed from nonimmunogenic materials while considering catheter hole configuration and size to deter ChP growth into the catheter holes and the lumen to prevent cellular catheter occlusion.
Keywords: BioGlide, Choroid plexus, Hydrocephalus, Immunoreactivity, Polyvinylpyrrolidone, Ventricular catheter obstruction
INTRODUCTION
Hydrocephalus is one of the leading causes of morbidity and mortality in infants, with ventricular shunting serving as the primary treatment.[
Many studies have sought to understand the biological basis of tissue infiltration and subsequent catheter obstruction. In 1981, Go et al. were one of the first groups to evaluate explanted, occluded catheters histologically. Their study found cellular detritus within the catheter as early as 1-month postimplantation.[
Given that ventricular catheters are susceptible to occlusion, researchers and engineers have sought to create occlusion-proof catheters. One prominent innovation is the use of polyvinylpyrrolidone (PVP), commonly known as BioGlide®, as a coating on ventricular catheters. PVP was initially created as a hydrophilic coating aimed at decreasing shunt failure by limiting bacterial colonization.[
CASE DESCRIPTION
The patient had a history of hydrocephalus secondary to Dandy-Walker malformation with a ventricular shunt subsequently placed at 8 months of age. The patient presented 8 years after their initial shunt placement with symptoms of shunt failure. The shunt was tapped percutaneously through needle aspiration, with no return of CSF. Imaging taken at the time of presentation demonstrated ventricular dilation compared with the baseline diminutive ventricular caliber [
During shunt exploration, there was no CSF flow from the PVP-coated ventricular catheter, and the proximal catheter was adherent within the ventricle. A pen-type endoscope was passed through the internal lumen of the ventricular catheter and internal debris was detached from its anchor point, freeing the catheter without removing the debris that was attached to it [
Pathologic findings and experimentation
The explanted PVP-coated ventricular catheter was analyzed in an experimental in vitro model, characterizing cellular occlusion and hydrodynamic properties by utilizing pressure and flow rate measurements. CSF flow through the catheter was conducted with a ventricular phantom model, syringe pump (Fusion 200-X, Chemyx Inc®), pressure sensor (px409-100GUSBH, Omega®), and flow rate sensor (SLF3S-0600F, Sensirion®) as previously reported.[
Figure 3:
Comparison of polyvinylpyrrolidone (PVP)-coated versus barium-impregnated obstructed catheter pressure and flow analysis. (a) Intracranial pressure-like graphic representation of the proximal catheters tested (PVP-coated vs. barium impregnated). Both catheters indicate malfunction as the intracranial-like pressure continued increasing while the pump was in space between dashed lines. (b) Graphic representation of catheter flow. No flow was found in either catheter, indicating total obstruction.
Figure 4:
Histological analysis of polyvinylpyrrolidone (PVP)-coated versus barium-impregnated obstructed catheters. (a) Histological preparation stained with hematoxylin and eosin (H&E) of a transverse slice of a barium-impregnated partially obstructed catheter (Magnification: 4x). (b) Magnification (10x) of the square in a, highlighting partial obstruction by a fibrinogen layer. (c) Immunofluorescence against 4',6-diamidino-2-phenylindole (DAPI, nuclear counterstain), fibrinogen, and Iba-1 (macrophage marker) in a parallel slice from b. The image shows high fibrinogen levels associated with the choroid plexus (ChP) (arrows) and, internally, the presence of macrophages (arrowhead). (d) Immunohistochemistry against transthyretin (TTR, choroid plexus marker) confirmed that ChP obstructed the catheter (arrow). (e-h) Histological preparations from an obstructed PVP-coated catheter with ChP similarly as in a-d. The images do not show a fibrinogen layer, as seen in the barium-impregnated catheter. Instead, low expression can be seen internally as part of the connective tissue (arrow). (h) Immunohistochemistry against TTR confirmed that ChP obstructed the polyvinylpyrrolidone-coated catheter (arrow). Arrowhead indicates few macrophages. Scale bars: a, d, e, and h; 500 µm; b, c, f, and g; 200 µm.
Analyses revealed that ChP had grown into the catheter holes and attached itself to the contralateral ChP in the lumen of the catheter [
DISCUSSION
There is an extensive history of shunt catheter surface modifications to mitigate shunt failure. Nulsen and Spitz introduced the first indwelling ventricular shunt system using rubberized catheters.[
In 1992, Gower et al. utilized expanded polytetrafluoroethylene (ePTFE) as a novel shunt catheter material.[
At present, histological research specific to explanted PVP-coated ventricular catheters is limited. Other studies with respect to failed shunts draw a parallel to the current findings. Sarkiss et al. conducted a histological analysis of 85 explanted catheters examining characteristics of occluded shunts as a function of time to revision.[
Hariharan et al. conducted a large histological analysis of explanted ventricular catheters, reporting histological findings on 44 devices.[
ChP has been found to play an emerging role in the previously described immune response. It is located in the ventricles of the brain and is highly vascularized tissue comprised of modified ependymal cells surrounding a central stroma and vasculature, responsible for the production of CSF and clearing of waste.[
In this case, we postulate that the lower distal pressure in the shunt system (i.e., siphoning) promoted the “pull” of ChP into the catheter. While the gross appearance of an occluded silicone catheter would be similar, our histopathological studies demonstrated a different mechanism of occlusion: ingrowth without cellular attachment in PVP-coated catheters and cellular attachment with ingrowth on the catheter walls and into the lumen in non-PVP-coated catheters. For future catheter design considerations, the modification of the catheter hole angle and irregularity of hole placement may aid in the prevention of the ingrowth of ChP into the catheter lumen.[
A novel catheter design to incorporate these modifications may be beneficial in understanding the CNS immunological response leading to silicone catheter occlusion. We propose that future designs should consider PVP-coated material and its potential to mitigate the initiation of an inflammatory response to the catheter surface.
This study is a single case report of one patient involving mechanisms of PVP-coated ventricular catheter obstruction and findings cannot be generalized for broader populations. However, these findings may promote potential new hypotheses and methodologies to be formed involving this field of research.
CONCLUSION
This case demonstrated the lack of immunoreactivity in a PVP-coated ventricular catheter. Considering previous studies of explanted shunt catheters, this finding adds to the ongoing evaluation of ventricular catheter coating materials.
While there are still significant obstacles when designing the optimal ventricular shunt catheter, this case report adds promising evidence to encourage catheter surface modification in future designs to mitigate a foreign-body immunological response and subsequent failure impacting patients.
Ethical approval
Institutional Review Board approval is not required.
Declaration of patient consent
The authors certify that they have obtained all appropriate patient consent.
Financial support and sponsorship
This project was supported by the LeVecke Family Foundation. Catheters were donated by Medtronic. We thank University Lab Partners at UC Irvine Research Park for providing the laboratory infrastructure necessary to complete this work.
Conflicts of interest
There are no conflicts of interest.
Use of artificial intelligence (AI)-assisted technology for manuscript preparation
The authors confirm that there was no use of artificial intelligence (AI)-assisted technology for assisting in the writing or editing of the manuscript and no images were manipulated using AI.
Disclaimer
The views and opinions expressed in this article are those of the authors and do not necessarily reflect the official policy or position of the Journal or its management. The information contained in this article should not be considered to be medical advice; patients should consult their own physicians for advice as to their specific medical needs.
References
1. Bayston R, Milner RD. Antimicrobial activity of silicone rubber used in hydrocephalus shunts, after impregnation with antimicrobial substances. J Clin Pathol. 1981. 34: 1057-62
2. Cagavi F, Akalan N, Celik H, Gur D, Guciz B. Effect of hydrophilic coating on microorganism colonization in silicone tubing. Acta Neurochir (Wien). 2004. 146: 603-10 discussion 609-10
3. Castaneyra-Ruiz L, Gonzalez-Marrero I, Hernandez-Abad LG, Carmona-Calero EM, Meyer G, Castaneyra-Perdomo A. A distal to proximal gradient of human choroid plexus development, with antagonistic expression of glut1 and AQP1 in mature cells vs. Calbindin and PCNA in proliferative cells. Front Neuroanat. 2016. 10: 87
4. Castaneyra-Ruiz L, Hernandez-Abad LG, Carmona-Calero EM, Castaneyra-Perdomo A, Gonzalez-Marrero I. AQP1 overexpression in the CSF of obstructive hydrocephalus and inversion of its polarity in the choroid plexus of a chiari malformation type II case. J Neuropathol Exp Neurol. 2019. 78: 641-7
5. Contardi M, Summa M, Picone P, Brancato OR, Di Carlo M, Bertorelli R. Evaluation of a multifunctional polyvinylpyrrolidone/hyaluronic acid-based bilayer film patch with anti-inflammatory properties as an enhancer of the wound healing process. Pharmaceutics. 2022. 14: 483
6. Gluski J, Zajciw P, Hariharan P, Morgan A, Morales DM, Jea A. Characterization of a multicenter pediatric-hydrocephalus shunt biobank. Fluids Barriers CNS. 2020. 17: 45
7. Go KG, Ebels EJ, van Woerden H. Experiences with recurring ventricular catheter obstructions. Clin Neurol Neurosurg. 1981. 83: 47-56
8. González-Marrero I, Giménez-Llort L, Johanson CE, Carmona-Calero EM, Castañeyra-Ruiz L, Brito-Armas JM. Choroid plexus dysfunction impairs beta-amyloid clearance in a triple transgenic mouse model of Alzheimer’s disease. Front Cell Neurosci. 2015. 9: 17
9. Gower DJ, Watson D, Harper D. e-PTFE ventricular shunt catheters. Neurosurgery. 1992. 31: 1132-5
10. Hakim S. Observations on the physiopathology of the CSF pulse and prevention of ventricular catheter obstruction in valve shunts. Dev Med Child Neurol Suppl. 1969. 20: 42-8
11. Hariharan P, Sondheimer J, Petroj A, Gluski J, Jea A, Whitehead WE. A multicenter retrospective study of heterogeneous tissue aggregates obstructing ventricular catheters explanted from patients with hydrocephalus. Fluids Barriers CNS. 2021. 18: 33
12. Harris CA, McAllister JP. What we should know about the cellular and tissue response causing catheter obstruction in the treatment of hydrocephalus. Neurosurgery. 2012. 70: 1589-601 discussion 1601-2
13. Harris CA, Morales DM, Arshad R, McAllister JP, Limbrick DD. Cerebrospinal fluid biomarkers of neuroinflammation in children with hydrocephalus and shunt malfunction. Fluids Barriers CNS. 2021. 18: 4
14. Ingraham FD, Alexander E, Matson DD. Polyethylene, a new synthetic plastic for use in surgery; Experimental applications in neurosurgery. J Am Med Assoc. 1947. 135: 82-7
15. Kraemer MR, Koueik J, Rebsamen S, Hsu DA, Salamat MS, Luo S. Overdrainage-related ependymal bands: A postulated cause of proximal shunt obstruction. J Neurosurg Pediatr. 2018. 22: 567-7
16. Lee S, Bristol RE, Preul MC, Chae J. Three-dimensionally printed microelectromechanical-system hydrogel valve for communicating hydrocephalus. ACS Sens. 2020. 5: 1398-404
17. McVerry B, Polasko A, Rao E, Haghniaz R, Chen D, He N. A readily scalable, clinically demonstrated, antibiofouling zwitterionic surface treatment for implantable medical devices. Adv Mater. 2022. 34: e2200254
18. Nulsen FE, Spitz EB. Treatment of hydrocephalus by direct shunt from ventricle to jugular vain. Surg Forum. 1951. p. 399-403
19. Panagopoulos D, Strantzalis G, Gavra M, Boviatsis E, Korfias S. The role of antisiphon devices in the prevention of central ventricular catheter obliteration for hydrocephalus: A 15-years institution’s experience retrospective analysis. Children (Basel). 2022. 9: 493
20. Riva-Cambrin J, Kestle JR, Holubkov R, Butler J, Kulkarni AV, Drake J. Risk factors for shunt malfunction in pediatric hydrocephalus: A multicenter prospective cohort study. J Neurosurg Pediatr. 2016. 17: 382-90
21. Sarkiss CA, Sarkar R, Yong W, Lazareff JA. Time dependent pattern of cellular characteristics causing ventriculoperitoneal shunt failure in children. Clin Neurol Neurosurg. 2014. 127: 30-2
22. Sekhar LN, Moossy J, Guthkelch AN. Malfunctioning ventriculoperitoneal shunts. Clinical and pathological features. J Neurosurg. 1982. 56: 411-6
23. Shannon CN, Acakpo-Satchivi L, Kirby RS, Franklin FA, Wellons JC. Ventriculoperitoneal shunt failure: An institutional review of 2-year survival rates. Childs Nerv Syst. 2012. 28: 2093-9
24. Sobolčiak P, Popelka A, Mičušík M, Sláviková M, Krupa I, Mosnáček J. Photoimmobilization of zwitterionic polymers on surfaces to reduce cell adhesion. J Colloid Interface Sci. 2017. 500: 294-303
25. Stone JJ, Walker CT, Jacobson M, Phillips V, Silberstein HJ. Revision rate of pediatric ventriculoperitoneal shunts after 15 years. J Neurosurg Pediatr. 2013. 11: 15-9
26. Strominger I, Elyahu Y, Berner O, Reckhow J, Mittal K, Nemirovsky A. The choroid plexus functions as a niche for T-cell stimulation within the central nervous system. Front Immunol. 2018. 9: 1066
27. Tully HM, Doherty D, Wainwright M. Mortality in pediatric hydrocephalus. Dev Med Child Neurol. 2022. 64: 112-7
28. Woo PY, Zhuang JT, Ho JM, Seto A, Wong HT, Chan KY. Intraventricular urokinase to treat a blocked ventriculoperitoneal shunt in a glioblastoma patient with leptomeningeal spread. Acta Neurochir (Wien). 2018. 160: 1073-77
29. Xu H, Huang Y, Jiao W, Sun W, Li R, Li J. Hydrogel-coated ventricular catheters for high-risk patients receiving ventricular peritoneum shunt. Medicine (Baltimore). 2016. 95: e4252