© 2010 Park P This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.How to cite this article: Park P, Foley KT, Cowan JA, Marca FL. Minimally invasive pedicle screw fixation utilizing O-arm fluoroscopy with computer-assisted navigation: Feasibility, technique, and preliminary results. Surg Neurol Int 25-Aug-2010;1:44
How to cite this URL: Park P, Foley KT, Cowan JA, Marca FL. Minimally invasive pedicle screw fixation utilizing O-arm fluoroscopy with computer-assisted navigation: Feasibility, technique, and preliminary results. Surg Neurol Int 25-Aug-2010;1:44. Available from: http://sni.wpengine.com/surgicalint_articles/minimally-invasive-pedicle-screw-fixation-utilizing-o-arm-fluoroscopy-with-computer-assisted-navigation-feasibility-technique-and-preliminary-results/
Abstract
Background:Pedicle screw misplacement is relatively common, with reported rates ranging up to 42%. Although computer-assisted image guidance (CaIG) has been shown to improve accuracy in open spinal surgery, its use in minimally invasive procedures has not been as well evaluated. We present our technique and review the results from a cohort of patients who underwent minimally invasive lumbar pedicle screw placement utilizing the O-arm imaging unit in conjunction with the StealthStation Treon System.
Methods:A retrospective review of patients who underwent minimally invasive pedicle screw fixation with CaIG was performed. Eleven consecutive patients were identified and all were included. Nine patients underwent a single-level transforaminal lumbar interbody fusion. Two patients underwent multi-level fusion. Inaccurate pedicle screw placement was determined by postoperative computed tomography (CT) and graded as 0–2, 2–4, 4–6, or 6–8 mm.
ResultsA total of 52 screws were placed. Forty screws were inserted in eight patients who had postoperative CT, and a misplacement rate of 7.5% was noted including one lateral and two medial breaches. All breaches were graded as 0–2 mm and were asymptomatic. In the remaining three patients, post-instrumentation O-arm imaging did not demonstrate pedicle screw misplacement.
ConclusionAlthough this initial study evaluates a relatively small number of patients, minimally invasive pedicle screw fixation utilizing the O-arm and StealthStation for CaIG appears to be safe and accurate.
Keywords: Image-guided surgery, minimally invasive procedures, O-arm, pedicle screw
INTRODUCTION
Pedicle screw misplacement is a potential complication of instrumented spinal fusion, which can lead to neurologic injury.[
METHODS
After getting approval from the institutional review board, a retrospective review of computerized medical records was performed. From December 2007 to April 2008, 11 consecutive patients underwent minimally invasive lumbar pedicle screw fixation utilizing the O-arm with Stealth navigation. All the patients were included in the study. Demographic information, surgical data, radiographic data, and complications were recorded.
Nine patients underwent a single-level minimally invasive transforaminal lumbar interbody fusion (MI-TLIF) [
Surgical technique
For the MI-TLIF, either the Sextant (Medtronic Sofamor Danek, Minneapolis, MN, USA) or the Viper (Depuy Spine, Raynham, MA, USA) system was used for pedicle screw fixation. Access for performing the decompression and interbody cage placement was obtained using the METRx tubular retraction system (Medtronic Sofamor Danek). The StealthStation Treon Guidance System with Synergy software (Medtronic) was used for CaIG, in conjunction with the O-arm fluoroscopic unit.
Each patient was positioned prone on a radiolucent Jackson frame with the StealthStation camera placed at the foot. The O-arm was then positioned so that the target spinal segment was centered in the field of view. To accomplish this, anterior-posterior (A/P) and lateral views were obtained in 2-D fluoroscopy mode. The O-arm was then moved toward the patient’s head to the “parked” position, which allowed access to the patient for the surgical procedure [
With the O-arm moved to the patient’s head, two paramedian 1-inch incisions overlying the target spinal segment were then made. The Stealth long drill guide (LDG) was then inserted through the fascia and used to locate the proper starting site and trajectory for pedicle screw placement. Positioning of the LDG was entirely based on image guidance without visualization of the spinal anatomy [Figures
Radiographic analysis
Postoperative imaging in eight patients consisted of a standard spine protocol CT scan with 2.5 mm slices. In the remaining three patients (4, 6, and 11), the O-arm was utilized to confirm adequate screw placement, so postoperative CT was not performed. Patient 6 did, however, undergo postoperative, non-spine protocol CT.
Accuracy of screw placement was recorded using a previously reported scale in which medial or lateral penetration of the cortex was graded as 0–2, 2–4, 4–6, or 6–8 mm.[
RESULTS
There were seven male patients (63.6%) in our study. The mean age for all patients was 44.0 years (range, 19–62 years). A total of 52 screws were placed. In the eight patients who had postoperative spine protocol CT imaging with sagittal and coronal reformats, 40 screws were placed. There were three (7.5%) instances out of the 40 screw insertions where the cortex was penetrated, consisting of two medial breaches and one lateral breach. All three cases were graded as 0–2 mm and none were symptomatic [
Twelve additional pedicle screws were placed in three patients who did not undergo postoperative spine protocol CT imaging. One patient had a non-spine protocol CT, which upon review, did not show evidence of screw misplacement. In the remaining two patients, the O-arm was used to check screw placement at the end of the procedure. Although not of the same diagnostic quality as a true CT scanner, the O-arm images obtained did not demonstrate significant medial or lateral screw misplacement [
DISCUSSION
Even with fluoroscopic assistance, inaccurate pedicle screw placement can be relatively common and can result in neurologic injury. Castro et al.[
In comparison to fluoroscopy-assisted screw placement, CaIG has been shown to achieve overall higher rates of accuracy. Merloz et al.[
Although many studies have now shown CaIG to increase accuracy in open spinal procedures, the use of CaIG in minimally invasive cases has not been as well evaluated. In minimally invasive procedures, exposure of the spine is typically very limited. Since early versions of CaIG relied on preoperative CT images of the spine registered to the patient intraoperatively by means of a point registration to the exposed spine, the lack of bony exposure limited the use of CaIG. However, a relatively recent innovation, the Siremobil IsoC 3-D fluoroscope (Siemens Medical Solutions, Erlangen, Germany), has made the use of CaIG in minimally invasive procedures both practical and feasible. Instead of relying on a preoperative CT image, the Iso-C can acquire multi-planar images intraoperatively with the patient positioned for surgery. After positioning of the StealthStation camera and placement of the reference arc, the Iso-C fluoroscope obtains successive images while rotating 190° around the patient. Approximately three adjacent vertebral levels can be imaged. This data are then transferred to the StealthStation navigation system, which performs an auto-registration process. Since there is no requirement for surgeon-derived point registration to the exposed spine, the use of Iso-C for minimally invasive pedicle screw fixation is possible. Villavicencio et al.
[
The O-arm fluoroscopic unit is a new imaging device that incorporates a flat panel detector and the X-ray source in a cylindrical bore. The cylindrical bore can be opened laterally to allow positioning around the operating room table. The O-arm can acquire standard fluoroscopic images as well as multi-planar images of the spine. Image acquisition in multi-planar mode involves successive images obtained while the X-ray source and detector rotate 360° around the patient. Similar to the Iso-C, data are automatically transferred to the StealthStation where auto-registration occurs for CaIG. Purported advantages of the O-arm compared to the Iso-C include improved multi-planar image quality due to the larger number of images obtained during the 360° acquisition, larger field of view, and robotic re-positioning of the O-arm to pre-programmed positions facilitating subsequent fluoroscopic views. One disadvantage is the expense of the O-arm fluoroscopic unit, which currently lists for approximately $700,000.00. In addition, the StealthStation Treon Guidance System costs approximately $250,000.00 [
This study represents our initial results of using the O-arm for minimally invasive pedicle screw insertion. The misplacement rate of 7.5% with no breach greater than 2 mm is comparable to previously reported rates of accuracy using CaIG in open spinal procedures. In addition, accurate screw placement was possible even with a significant scoliosis and over four adjacent vertebral segments with a single registration due to the large field of view.
Study limitations
Although O-arm fluoroscopy with CaIG appears to allow accurate placement of minimally invasive pedicle screws, this study is retrospective and involves a relatively small number of patients. In addition, while all patients underwent a minimally invasive approach in which screws were inserted without direct visualization of the pedicle screw entry sites, the population of patients studied was not uniform, with two patients undergoing a hybrid procedure. To fully validate safety and efficacy, a larger prospective comparative study is required.
CONCLUSION
The O-arm is a new intraoperative imaging system that is capable of acquiring both fluoroscopic and multi-planar images which can be seamlessly integrated with the StealthStation for CaIG. Minimally invasive pedicle screw insertion utilizing the O-arm and CaIG appears to be safe and accurate.
References
1. Castro WH, Halm H, Jerosch J, Malms J, Steinbeck J, Blasius S. Accuracy of pedicle screw placement in lumbar vertebrae. Spine. 1996. 21: 1320-4
2. Esses SI, Sachs BL, Dreyzin V. Complications associated with the technique of pedicle screw fixation: A selected survey of ABS members. Spine. 1993. 18: 2231-9
3. Gertzbein SD, Robbins SE. Accuracy of pedicular screw placement in vivo. Spine. 1990. 15: 11-4
4. Holly LT, Schwender JD, Rouben DP, Foley KT. Minimally invasive transforaminal lumbar interbody fusion: Indications, technique, and complications. Neurosurg Focus. 2006. 20: E6-
5. Laine T, Lund T, Ylikoski M, Lohikoski J, Schlenzka D. Accuracy of pedicle screw insertion with and without computer assistance: A randomised controlled clinical study in 100 consecutive patients. Eur Spine J. 2000. 9: 235-40
6. Laine T, Makitalo K, Schlenzka D, Tallroth K, Poussa M, Alho A. Accuracy of pedicle screw insertion: A prospective CT study in 30 low back patients. Eur Spine J. 1997. 6: 402-5
7. Merloz P, Tonetti J, Pittet L, Coulomb M, Lavallee S, Sautot P. Pedicle screw placement using image guided techniques. Clin Orthop Relat Res. 1998. 354: 39-48
8. Villavicencio AT, Burneikiene S, Bulsara KR, Thramann JJ. Utility of computerized isocentric fluoroscopy for minimally invasive spinal surgical techniques. J Spinal Disord Tech. 2005. 18: 369-75