DOI:10.4103/2152-7806.139664
Copyright: © 2014 Xu R. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.How to cite this article: Xu R, Bydon M, Macki M, Belkoff SM, Langdale ER, McGovern K, Wolinsky J, Gokalsan ZL, Bydon A. Biomechanical impact of C2 pedicle screw length in an atlantoaxial fusion construct. Surg Neurol Int 28-Aug-2014;5:
How to cite this URL: Xu R, Bydon M, Macki M, Belkoff SM, Langdale ER, McGovern K, Wolinsky J, Gokalsan ZL, Bydon A. Biomechanical impact of C2 pedicle screw length in an atlantoaxial fusion construct. Surg Neurol Int 28-Aug-2014;5:. Available from: http://sni.wpengine.com/surgicalint_articles/effects-of-teriparatide-on-lung-function-and-pain-relief-in-women-with-multiple-osteoporotic-vertebral-compression-fractures/
Abstract
Background:Posterior, atlantoaxial (AA) fusions of the cervical spine may include either standard (26 mm) or short (16 mm) C2 pedicle screws. This manuscript focused on an in vitro biomechanical comparison of standard versus short C2 pedicle screws to perform posterior C1-C2 AA fusions.
Methods:Twelve human cadaveric spines underwent C1 lateral mass screw and standard C2 pedicle screw (n = 6) versus short C2 pedicle screw (n = 6) fixation. Six additional controls were not instrumented. The peak torque, peak rotational interval, and peak stiffness of the constructs were analyzed to failure levels.
Results:The peak torque to construct failure was not statistically significantly different among the control spine (12.2 Nm), short pedicle fixation (15.5 Nm), or the standard pedicle fixation (11.6 Nm), P = 0.79. While the angle at the peak rotation statistically significantly differed between the control specimens (47.7° of relative motion) and the overall instrumented specimens (P P = 0.39). Similarly, although the average stiffness was statistically significantly lower in control group (0.026 Nm/degree) versus the overall instrumented specimens (P = 0.001), the standard C2 pedicle screws (2.54 Nm/degree) did not differ from the short C2 pedicle screws
Conclusions:Both standard and short C2 pedicle screws allow for equally rigid fixation of C1 lateral mass-C2 AA fusions. Usage of a short C2 pedicle screw may be an acceptable method of stabilization in carefully selected patient populations.
Keywords: Atlantoaxial, biomechanics, C1, C2, pars, pedicle
INTRODUCTION
Spinal C2 screw fixation utilizing a shorter versus a longer screw may minimize neurovascular injuries.[
MATERIALS AND METHODS
Fresh frozen human cadaveric spines (occiput-T6) were obtained from the Maryland State Anatomy Board. All spines underwent radiographic inspection via anterior-posterior and lateral X-ray in addition to Dual Energy X-Ray Absorptiometry (DEXA) scanning. Apparent bone mineral density and absence of osteoporosis (t scores greater than -2.5) were confirmed with Dual Energy X-ray Absorptiometry (Discovery QDR Series, Hologic Inc., Bedford, Massachusetts). Any spines demonstrating fractures, tumors, abnormal radiolucency, or other pathologies that may compromise the structural integrity of the specimens were discarded. All spines were stored in sealed plastic bags at -20°C until 1 day before testing, at which time they were allowed to thaw overnight to room temperature [
Figure 1
(a) All spines were stored at -20°C in sealed plastic bags until one day before testing, at which time they were thawed overnight to room temperature. (b) C1 lateral mass-standard C2 pedicle screw (26 mm) instrumentation. (c) C1 lateral mass-short C2 pedicle screw instrumentation. (d) Potting of spines with C1-C2 constructs
C1-C2 Instrumentation
Posterior C1-C2 cantilever fusion constructs (without hemilaminotomy) followed the technique described by Harms and Melcher.[
Assessment of spinal rigidity
The superior and inferior aspects of the specimens were potted in polyvinyl chloride pipe using polymethylmethacrylate bone cement (FastTray, Bosworth Co., Skokie, Illinois) [
Figure 2
(a) Potting of the inferior portion of the spine. (b) Fixation into the materials testing machine (MTS Bionix 858 Test System, MTS, Eden Prairie, Minnesota). (c) Positioning of reflective marker triads (SMART motion capture system, eMotion, Padova, Italy). (d) Testing to failure of the instrumented spines
The weight of the caudal fixture provided an axial load equivalent to 50N. Theses caudal forces replicated the degree of flexion (approximately 11°) at the AA joint.[
Statistical analysis
The objective in assessing spinal rigidity was primarily focused on peak torque; that is, rotational movements until failure. In a seminal publication on biomechanical evaluations of C1-C2 posterior constructs, Sim et al. determined that the mean range of motion for an intact spine was 47.5° ± 9.6.[
Following the biomechanical study, crude data was described with summary statistics. We used a generalized linear latent and mixed regression model in STATA 10 (StataCorp LP, College Station, TX) to analyze the effect of covariates (noninstrumented, short C2 pedicle screw, and standard C2 pedicle screw) on failure (peak) torque, stiffness, and angle at peak torque to generate 95% confidence intervals. We also analyzed our data via analysis of variance (ANOVA) with Tukey post-hoc analysis, wherein significant effect was defined as P < 0.05.
RESULTS
The peak torque to construct failure was not statistically significantly different among the control spine, short pedicle fixation, or the standard pedicle fixation, P = 0.79. The mean peak torque (95% confidence interval) to failure was 12.2 Nm (5.7-18.7 Nm) in the control cohort, 15.5 Nm (7.6-23.4 Nm) in the short pedicle screw cohort, and 11.6 Nm (9.8-13.4 Nm) in the standard pedicle screw cohort [
However, the angle at which the peak occurred did differ significantly. Failure occurred at an average of 47.7° of relative rotation (36.459°) in the control specimens, which was significantly greater than the instrumented specimens overall, P < 0.001: 20.7° of relative rotation (13.7°27.8°) in the short C2 pedicle screw specimens and 13.7° of relative rotation (7.1-20.2°) in the standard C2 pedicle screw specimens. Although the short C2 pedicle screw instrumentation afforded spines a higher degree of rotational range of motion, the treatments were not significantly different from each other, P = 0.39.
Although standard C2 pedicle screws afforded spines a higher average stiffness than short C2 pedicle screws, this was not statistically significant, P = 0.30. The measured stiffness was 0.026 Nm/degree (0.014–0.038 Nm/degree) for the control group, which was significantly less than the treated specimens, P = 0.001: 1.69 Nm/degree (1.08-2.30 Nm/degree) in the short pedicle screw group and 2.54 Nm/degree (1.43-3.66 Nm/degree) in the standard pedicle screw group.
DISCUSSION
As the placement of C2 pedicle screws is technically challenging due to the proximity of the vertebral artery laterally and the spinal canal medially, some spine surgeons opt for the placement of short pedicle screws.[
Our results demonstrate that the short C2 pedicle screw was able to achieve comparable rigidity to the standard C2 pedicle screw in a test to failure of a C1-C2 segmental screw-rod fusion system. The short C2 pedicle screw was able to significantly increase the angle at peak torque and maximum stiffness compared to noninstrumented controls. Compared with the standard C2 pedicle screw, the short C2 pedicle screw exhibited slightly higher rotational degrees of freedom and a lower degree of stiffness to failure, although these comparisons were not statistically significant. Taken together, our results suggest that short C2 pedicle screws afford similar levels of postinstrumentation rigidity compared with standard C2 pedicle screws.
Literature review
Although no study in the literature has directly examined the effect of C2 pedicle screw length on a biomechanical test to failure model, Sim et al. compared a short C2 pedicle screw (14-16 mm) to a standard long C2 pedicle screw (24-28 mm) in a C1-C2 AA fixation study.[
Clinically, our findings are also supported by other studies that show excellent results utilizing the short C2 pedicle screws and C2 pars screw, both of which allow for C1-C2 fixation without a standard length pedicle screw. Nitising et al. demonstrated no permanent complications in 10 patients undergoing C1 lateral mass-C2 pars screw instrumentation; in all cases, rigid fixation was confirmed on postoperative radiographs and maintained on follow-up radiographs.[
Limitations
As an in vitro biomechanical model, our results cannot fully replicate the in vivo fusion process replete with paraspinal muscles and endogenous bone turnover. Therefore, this study does not provide incontrovertible proof of long-term stability. Nonetheless, our biomechanical data provides mechanistic insight into the clinical observations by Mummaneni et al. and may serve to enhance our understanding of the value of the short C2 pedicle screw. Our biomechanical study was also limited regarding testing to failure; fatigue and quasistatic testing may provide additional data with postfatigue comparative analyses and force displacement in multiple planes, respectively.
CONCLUSIONS
This study showed that a shorter C2 pedicle screw provided equally rigid fixation of a C1 lateral mass-C2 segmental fusion construct in a biomechanical test to failure model. The short C2 pedicle screws (coupled with C1 lateral mass screws) may, therefore, serve as a biomechanically sound stabilizing method of fusion of the AA joint. Additional prospective studies in patient populations are warranted in order to corroborate these biomechanical findings.
References
1. Bransford RJ, Russo AJ, Freeborn M, Nguyen QT, Lee MJ, Chapman JR. Posterior C2 instrumentation: Accuracy and complications associated with four techniques. Spine (Phila Pa 1976). 2011. 36: E936-43
2. Elgafy H, Potluri T, Goel VK, Foster S, Faizan A, Kulkarni N. Biomechanical analysis comparing three C1-C2 transarticular screw salvaging fixation techniques. Spine (Phila Pa 1976). 2010. 35: 378-85
3. Grob D, Crisco JJ, Panjabi MM, Wang P, Dvorak J. Biomechanical evaluation of four different posterior atlantoaxial fixation techniques. Spine (Phila Pa 1976). 1992. 17: 480-90
4. Harms J, Melcher RP. Posterior C1-C2 fusion with polyaxial screw and rod fixation. Spine (Phila Pa 1976). 2001. 26: 2467-71
5. Lehman RA, Dmitriev AE, Helgeson MD, Sasso RC, Kuklo TR, Riew KD. Salvage of C2 pedicle and pars screws using the intralaminar technique: A biomechanical analysis. Spine (Phila Pa 1976). 2008. 33: 960-5
6. Mummaneni PV, Lu DC, Dhall SS, Mummaneni VP, Chou D. C1 lateral mass fixation: A comparison of constructs. Neurosurgery. 2010. 66: 153-60
7. Nitising A, Jetjumnong C, Tisavipat N, Nantaaree S. Posterior C1-C2 fusion using C1 lateral mass and C2 pars screw with rod fixation: Techniques and outcomes. J Med Assoc Thai. 2011. 94: 794-800
8. Panjabi M, Dvorak J, Duranceau J, Yamamoto I, Gerber M, Rauschning W. Three-dimensional movements of the upper cervical spine. Spine (Phila Pa 1976). 1988. 13: 726-30
9. Sim HB, Lee JW, Park JT, Mindea SA, Lim J, Park J. Biomechanical evaluations of various c1-c2 posterior fixation techniques. Spine (Phila Pa 1976). 2011. 36: E401-7